[scheme design] scheme of household oximeter based on single chip microcomputer

2022-06-11 09:09:00 Xicheng micro technology program development

   The oximeter is used to measure pulse rate 、 Instruments and equipment for blood oxygen concentration and perfusion index , Blood oxygen concentration is one of the important basic data in clinical medicine . If there are conditions , People should always monitor their oxygen saturation , It is more beneficial to your health . A scheme of household oximeter based on single chip microcomputer is introduced , The following is a detailed knowledge description .
  
   One 、 Measuring principle of oximeter
  
   The oximeter is based on the reduction of hemoglobin (Hb)、 Oxygenated hemoglobin (HbO2) The absorption spectrum characteristics in the red and near-infrared regions are the theoretical basis , Application LambertBeer Law establishes empirical formulas for data processing . The instrument adopts photoelectric blood oxygen detection technology combined with volume pulse tracing technology , Two beams of light with different wavelengths irradiate the fingertip of the human body through the perspective clip finger sensor, and the measurement signal is obtained by the photosensitive element , The obtained information is processed by the electronic circuit and microprocessor, and the measured results are displayed on the display screen .
  
   Two 、 Technical characteristics of the oximeter
  
  1、 Main performance of the oximeter
  
  A. The blood oxygen saturation value is displayed
  
  B. Pulse rate value 、 Bar graph display
  
  C. Pulse waveform display
  
  D. Battery level indication
  
  E. Automatic shutdown function : After the tested finger is detached 8s Internal automatic shutdown
  
  F. You can change the display direction , Depending on the display mode
  
  2、 Main parameters of the oximeter
  
  A. Blood oxygen saturation measurement range :35%~100% error : When the measurement range of blood oxygen saturation is 70%~100%, The allowable absolute error is ±2%; Less than 70% No definition .
  
  B. Pulse rate measurement range :30bpm( Time /min)~250bpm( Time /min) error :±2bpm or ±2% Take a big value .
  
  C. Display resolution : Blood oxygen saturation is 1%, Pulse rate is 1bpm.
  
  D. The error of weak perfusion is 0.3% when , The oximeter can correctly display oxygen saturation and pulse rate , The error of blood oxygen saturation is ±3%; Pulse rate error is ±2bpm or ±2% Take a big value .( remarks : This indicator only refers to high-end oximeters , Weak perfusion degree of low-end oximeter >=1)
  
  E. The anti ambient light interference ability compares the blood oxygen measurement value under indoor natural light and existing lighting sources with the measurement value under dark room conditions , The deviation is less than ±1%.
  
  F. Working voltage :2.6VDC~3.0VDC It shall work normally .
  
  G. Power waste :<30mA
  
  3、 Features of oximeter products
  
   Easy to use .
  
   High precision , Low power consumption .
  
   Small volume , Light weight , be easy to carry about .