当前位置:网站首页>Influence of oscilloscope probe on measurement bandwidth
Influence of oscilloscope probe on measurement bandwidth
2022-07-05 20:56:00 【PRBTEK】
Bandwidth is defined as the measurement system on the amplitude and frequency diagram is lower than the reference level 3dB The point of , Pictured 4.7 Shown , This shows that 3dB Response curve of point .
4.7. Bandwidth is defined as the decrease in amplitude in the response curve -3dB The frequency of
It is necessary to point out , The measuring system has a low amplitude in the rated bandwidth 3dB. This means that you can predict the frequency at the bandwidth limit , Amplitude measurement will have 30% The error of the .
Usually , Users will not use the oscilloscope with the full bandwidth limit . however , If amplitude accuracy is critical , Prepare to reduce the rated bandwidth of the oscilloscope accordingly .
for example , Consider the following figure 4.8 The expansion diagram of bandwidth roll off shown in . The horizontal scale in this figure shows that the result is better than 30% The rating reduction factor required for amplitude accuracy . If there is no rating reduction factor ( The coefficient is 1.0),100MHz Oscilloscope in 100MHz The amplitude error will be as high as 30%. If you want the amplitude measurement to fall on 3% Within the scope of , The bandwidth of this oscilloscope must be 0.3 The coefficient drops to 30MHz. When the frequency exceeds 30MHz when , The amplitude error will exceed 3%.
4.8. Rated bandwidth drop curve
The above example indicates the overall rule of thumb for oscilloscope selection . Yes 3% Amplitude measurement within , The specified bandwidth should be selected to be higher than the highest frequency waveform measured 3-5 Times the oscilloscope .
When the rising time or falling time is the main indicator , The following formula can be used to calculate the bandwidth of the oscilloscope (BW) The index is converted into rise time
indicators :
Tr≈0.35/BW
Or for convenience :
Tr(ns)≈350/BW(MHz)
Same as bandwidth , The rise time should be chosen to be faster than the estimated fastest rise time measured 3-5 Times the oscilloscope .( It should be pointed out that , The above bandwidth to rise time conversion assumes that the response of the oscilloscope is Gaussian roll off . Most oscilloscopes are designed for Gaussian roll off response .)
Oscilloscope probe bandwidth
Like other circuits , All oscilloscopes have bandwidth limits . Besides , Same as oscilloscope , The performance of the oscilloscope probe generally depends on the bandwidth . therefore , The bandwidth is 100MHz Your oscilloscope is 100MHz The amplitude response at point is lower than 3dB.
Allied , The oscilloscope probe bandwidth can also be expressed by the same formula used by the oscilloscope (Tr≈0.35/BW). Besides , For active probes , The following formula can be used to combine the rise time of oscilloscope and probe , Obtain an approximate probe / Rise time of oscilloscope system :
Trsystem2≈Trprobe2+Trscope2
For passive probes , This relationship is more complicated , The above formula should not be used .
Generally speaking , The probe bandwidth should always equal or exceed the bandwidth of the oscilloscope to be used . If the probe bandwidth used is too low , It will limit the oscilloscope to achieve all measurement functions . chart 4.9 Further illustrates this point , It shows the same pulse jump measured by three probes with different bandwidths .
4.9. Influence on the rise time of three different oscilloscope probes :(a)400MHz,10X Probe ;(b)100MHz,10X Probe ;(c)10MHz,1X Probe . All measurements use the same 400MHz Oscilloscope complete
Pictured 4.9a Shown , The first measurement is using matching 400MHz The combination of oscilloscope and probe . The probe used is 10 Megohm resistance and 14.1pF Capacitive 10X Probe . Be careful , The measured pulse rise time is 4.63ns. This falls well on 400MHz Oscilloscope / Probe combination 875ps Rise time range .
Now let's see how to use the same oscilloscope 、 Use 10X,100MHz What happens when the probe measures the same pulse , Pictured 4.9b Shown , Now the measured rise time is 5.97ns. This is more than previously measured 4.63ns Improved near 30%!
According to expectations , When using a probe with low bandwidth , The observed pulse rise time will become longer . Extreme examples are shown in the figure 4.9c Shown , Which is used on the same pulse 1X,10MHz Probe . here , The rise time has changed from the original 4.63ns Down to 27ns.
chart 4.9 The main conclusion is : Not any probe can do this !
To achieve the maximum performance of any oscilloscope , It is also a performance that you spend money on , Be sure to use the probe recommended by the manufacturer .
Bandwidth to oscilloscope probe tip
Generally speaking , It shall be possible to solve the probe bandwidth and the resulting probe according to the manufacturer's following specifications and recommendations / Oscilloscope system bandwidth . for example , Tektronix specifies the bandwidth of the probe working within the specified limit . These limits include global distortion 、 Rise time and scanning bandwidth .
in addition , When used with a compatible oscilloscope , Tektronix probe extends the bandwidth of the oscilloscope to the probe tip . for example , In compatible 100MHz When the oscilloscope is used , teck 100MHz The probe provides 100MHz Performance of (-3dB).
chart 4.10 The equivalent circuit in describes the industry recognized test setup used to verify the bandwidth to the probe tip . The specified signal source impedance of the test signal source is 50 ohm , Terminate at 50 Ohmic resistance , Lead to equivalent 25 Ohm source terminal impedance . Besides , The probe must use the probe head to BNC An adapter or equivalent device is connected to the signal source . The latter requirement of probe connection ensures the shortest grounding path .
4.10. The equivalent circuit used to test the bandwidth to the probe tip . Yes 100MHz System , The displayed rise time should be 3.5ns Or faster
When using the test setup described above ,100MHz Oscilloscope / The probe combination should result in the observed rise time <3.5ns. According to the bandwidth discussed above / Rise time relationship (Tr≈0.35/BW), this 3.5ns Rise time and 100MHz Bandwidth correspondence . For general purpose oscilloscope including standard matching probe , Most manufacturers promise 、 The declared oscilloscope bandwidth is provided on the probe tip .
however , Remember that the bandwidth on the probe tip depends on the figure 4.10 Test Method in . Because the actual environmental signal is rarely from 25 From ohmic signal source , Therefore, in the actual environment, the expected response and bandwidth should be lower than the optimal level to a certain extent , This is especially true when measuring higher impedance circuits .
Ground wire influence
When measuring the reference grounding , There must be two connections to the circuit or device under test . A connection is made through the probe , The probe senses the voltage of other measured parameters . Another necessary connection is to return to ground through the oscilloscope , Connect it back to the circuit under test . To complete the measurement of current path , It is necessary to realize the grounding circuit .
When the circuit under test and the oscilloscope are inserted into the same power socket circuit , The common potential of the power circuit provides a grounding loop . The signal circuit grounded through the power supply is generally indirect , The length is very long . result , You can't rely on this signal loop like a clean low inductance grounding loop .
Generally speaking , When measuring with any type of oscilloscope , The shortest grounding path should be used . The final grounding system is an online ECB( Etch circuit board ) To the probe head adapter , Pictured 4.11 Shown .ECB The adapter allows the probe tip to be directly inserted into the circuit test point , The outer barrel of the adapter should form a direct contact with the grounding ring of the probe head 、 Short ground contact . Measure the critical amplitude and timing , It is suggested that the established test points in the circuit board design include ECB/ Probe head adapter . This not only clearly indicates the location of the test point , And it ensures that the test points can be best connected , Realize the most reliable oscilloscope measurement .
4.11.ECB To the oscilloscope probe head adapter
边栏推荐
- The development of research tourism practical education helps the development of cultural tourism industry
- Duchefa丨P1001植物琼脂中英文说明书
- AITM2-0002 12s或60s垂直燃烧试验
- Duchefa p1001 plant agar Chinese and English instructions
- poj 3414 Pots (bfs+线索)
- Pytorch实战——MNIST数据集手写数字识别
- Open source SPL eliminates tens of thousands of database intermediate tables
- Duchefa low melting point agarose PPC Chinese and English instructions
- 线程池的使用
- 大二下个人发展小结
猜你喜欢
How to make ERP inventory accounts of chemical enterprises more accurate
台风来袭!建筑工地该如何防范台风!
Wanglaoji pharmaceutical's public welfare activity of "caring for the most lovely people under the scorching sun" was launched in Nanjing
解读协作型机器人的日常应用功能
Chemical properties and application instructions of prosci Lag3 antibody
Return to blowing marshland -- travel notes of zhailidong, founder of duanzhitang
Abbkine trakine F-actin Staining Kit (green fluorescence) scheme
10000+ 代码库、3000+ 研发人员大型保险集团的研发效能提升实践
Phpstudy Xiaopi's MySQL Click to start and quickly flash back. It has been solved
Open source SPL eliminates tens of thousands of database intermediate tables
随机推荐
sql系列(基础)-第二章 限制和排序数据
Abnova fluorescent dye 620-m streptavidin scheme
王老吉药业“关爱烈日下最可爱的人”公益活动在南京启动
Analyze the knowledge transfer and sharing spirit of maker Education
Monorepo management methodology and dependency security
五层网络协议
Duchefa low melting point agarose PPC Chinese and English instructions
bazel是否有学习的必要
Material design component - use bottomsheet to show extended content (II)
systemd-resolved 开启 debug 日志
Abbkine trakine F-actin Staining Kit (green fluorescence) scheme
CareerCup它1.8 串移包括问题
Graph embedding learning notes
清除app data以及获取图标
10000+ 代码库、3000+ 研发人员大型保险集团的研发效能提升实践
驱动壳美国测试UL 2043 符合要求有哪些?
Prosci LAG-3 recombinant protein specification
Duchefa丨D5124 MD5A 培养基中英文说明书
LeetCode: Distinct Subsequences [115]
Abnova maxpab mouse derived polyclonal antibody solution