当前位置:网站首页>Calibration of vector network analyzer (vector network)
Calibration of vector network analyzer (vector network)
2022-07-28 18:26:00 【Welcome pine 88】
One 、 error analysis
The errors in the testing process of network analyzer are mainly divided into three categories : System error 、 Random error 、 Drift error .
1、 The system error is caused by the unsatisfactory internal test device of the instrument , It is predictable and repeatable . Because it does not change with time , So it can be described quantitatively . The system error can be eliminated by calibration during the test .
2、 Random errors are unpredictable , Because it exists in random form , Will change over time , Therefore, it cannot be eliminated by calibration . The main source of random error is : Instrument internal noise ( Such as : Excitation source phase noise 、 Sampling noise 、 If receiver Lo noise ), The switch action repeatability and connector repeatability of the instrument also belong to random error .
3、 Drift error is the performance drift of the test device after the instrument is calibrated . Drift error is mainly caused by temperature change , Can be eliminated by further calibration . The length of time that the instrument can maintain stable accuracy after calibration depends on the drift speed of the instrument in the test environment .
Two 、 Specific analysis of systematic error
Since the calibration of network analyzer can eliminate the systematic error in the test process . The following is a specific analysis of the systematic error of the network analyzer in the reflection characteristic test process .
chart 2-1 The error content of network analyzer when testing the reflection characteristics of devices
1、 Frequency response error
The network analyzer works under the frequency scanning state , Whether it is the power divider inside the instrument , Directional coupler , Or external adapters and test cables , Within the working frequency band, its characteristics will change with frequency . The test error caused by these frequency response characteristics is called “ Frequency response error ”, Also known as “ Tracking error ”. Frequency response error includes reflection tracking error in reflection characteristic test and transmission tracking error in transmission characteristic test .
2、 Directional error
The error caused by the finite directivity of the directional coupler is the directivity error , The leakage signal caused by directional error will be superimposed on the real reflected signal , Cause test error . When the port matching performance of the tested part is good , The directional error has a great influence on the test .
3、 Port mismatch error
During the reflection index test , The reflected signal returns to the instrument port through the transmission path , There will be a mismatch between the impedance of the instrument port and the transmission line , This mismatch will cause secondary incidence of the signal , Finally, the multiple incidence of the signal in the transmission path , Accordingly, multiple reflections are formed , This error is called source mismatch error . The worse the matching performance of the tested part , The more obvious the influence of this error on the test .
Again , The transmission signal output by the device under test will also be reflected due to the impedance mismatch at the receiving end , The signal will be superimposed on the real reflected signal through the reverse transmission of the tested part . Thus, the load mismatch error is formed . If the reverse transmission isolation performance of the tested part is poor , The influence of load mismatch error is great .
4、 Isolation error
Inside the network analyzer R1、 A、 B The receiver should reflect the test input separately 、 Reflect and transmit signals . But there will be signal crosstalk between these receivers , For devices under test with high isolation ( switch 、 Isolators 、 Wide range attenuator ) Test of . The effect of this error is obvious .
Upper figure 2-1 in , There are common problems in the forward reflection characteristic test 6 Item error , There is symmetry in reverse testing 6 Item test error , Therefore, there are two port device tests 12 Item error .
3、 ... and 、 Calibration principle and method
The calibration principle is to measure the calibration device with known parameters , Store these measurement results in the memory of the analyzer , Use these data to calculate the error model . then , Use the error model to remove the influence of system error from the subsequent measurement .
The calibration process is the process of determining the instrument system error by testing the calibration pieces . According to the different calibration pieces , Calibration methods can be divided into mechanical calibration and electronic calibration . According to the difference of eliminating the error term , Mechanical calibration can be divided into frequency response calibration and vector calibration . Vector calibration can be divided into single port 、 Dual port 、 Multi port and TRL calibration .
Number of calibration pieces for each calibration method 、 The number of tests and the number of error elimination items are different . The accuracy of calibration from high to low is : TRL calibration 、 Electronic calibration 、 Vector calibration 、 Frequency response calibration .
3.1 Frequency response calibration
Frequency response calibration (Response) Only test 1 Calibration pieces , Use only 1 Calibration test operation .
It is a total reflection calibrator during reflection test , Short circuit calibrators can be used (Short) Or open circuit calibrator (Open). Generally, use terminal short circuit (Short) Closer to the ideal total reflection state .
During transmission test , Use straight through calibrators (Through).
Frequency response calibration is relatively simple , Low precision , Only frequency response errors are eliminated . The frequency response calibration process is equivalent to the test normalization process . That is, first store the test results in the memory to get the reference line , Then compare with the test results of the tested piece . In this way, the influence of systematic errors in the reference line can be eliminated .
3.2 Vector calibration
Vector calibration requires the network analyzer to have the ability to test amplitude and phase , Simultaneous equations are needed in the process of calculating the error term . The vector calibration process is more complex , It is required to test multiple standard parts , Thus, more error terms can be eliminated , Ensure that the instrument has higher test accuracy .
3.2.1 Single port vector calibration
Single port calibration (1-Port Cal) Need to use 3 Calibration pieces ( Open、 Short、 Load), Conduct 3 Calibration test operation .
When the calibration port is the port of the instrument 1 when , be called S11 Single port calibration ;
When the calibration port is the port of the instrument 2 when , be called S22 Single port calibration .
Single port calibration can eliminate the 3 Item systematic error ( Directional error 、 Source mismatch error 、 Reflection tracking error ).
3.2.2 Dual port vector calibration
Dual port calibration (2-Port Cal) Need to use 4 Calibration pieces ( Open、 Short、 Load、 Through), Conduct 7 Calibration test
Trial operation . The dual port isolation calibration is only used to test high isolation ( Isolators 、 switch )、 Large dynamic range ( filter ) Only when the device
be used .
When the network analyzer is used for the transmission performance test of the tested device , It is necessary to calibrate the test port and transmission connection line of the network analyzer . Dual port calibration eliminates all of the two test ports 12 Item systematic error .
3.2.3 Multi port vector calibration
Multi port calibration (3-Port Cal or 4-Port Cal) It is a pairwise combination of dual port calibration , Therefore, it is also necessary to use 4 Calibration pieces , But the calibration test operation will increase .
3.2.4 TRL calibration
TRL Calibration also belongs to vector calibration , Only for dual port and multi port calibration , However, it is different from the calibration devices and test methods used in the traditional dual port vector calibration described above .
It is not easy to accurately determine the parameters of calibration parts in traditional mechanical calibration , Because there will be parasitic inductance in the short circuit , There will be parasitic capacitance in the open circuit . and TRL The calibration uses transmission line devices , Its parameters are easier to establish , And the calibration accuracy is not completely determined by the calibration piece .
TRL Use three kinds of calibration pieces : Straight through calibration piece (Through)、 Reflection calibration piece (Reflect)、 Transmission line calibration piece (Line).
chart 3-1 TRL calibration
3.3 Electronic calibration
Agilent ENA The series network analyzer can be calibrated with traditional mechanical calibrators , Electronic calibrators can also be used ( E-Cal). Agilent ENA The communication control between the network analyzer and the electronic calibrator adopts USB Interface . Compared with mechanical calibration pieces , The electronic calibration piece has the following characteristics :
1、 The calibration process is simple , The electronic calibrator only needs to be connected to the vector network once , The test items required for dual port calibration can be completed . There is no need to connect the calibration pieces many times .
2、 Fast calibration , Using electronic calibration to complete dual port calibration takes only a few seconds , The efficiency of the whole testing process is greatly improved .
3、 There are few uncertainties in the calibration process , Because there is no need to connect multiple times , Therefore, the probability of electronic calibration affected by misoperation will be reduced .
4、 Support mixed port calibration , Agilent Electronic calibrators in the form of mixed ports are available , It can ensure the accuracy of testing many non inserted devices .
chart 3-2 Electronic calibration
Four 、 Calibration method
The operation of calibration mainly includes the following steps :
1、 First , Select the port to be calibrated and tested according to the tested devices and test items ( Single port 、 Dual port );
2、 secondly , Determine the model of calibration piece ( Model of electronic calibration or specific model of mechanical calibration piece ) ;
3、 Again , Determine the way of calibration ( Electronics 、 Frequency response 、 Vectors, etc ) ;
4、 then , Enter the specific calibration test process . Such as , If the machine is calibrated , After connecting the Colonel approval , Then click the option corresponding to the network partition ;
5、 Finally, the storage and call of test status and calibration parameters .
边栏推荐
- Syntax error: non declaration statement outside function bodygo and syntax error: unexpected {, expect
- USB Type-C 详解
- LeetCode79题 方法一:深度搜索
- 7-day sales of video number exceed one million
- Andorid: pit encountered by zxing.demo QR code scanning framework
- Mongodb create index
- 高温天气户外活动有讲究!市民盛夏健身安全指引来了
- Import the database backup of MySQL 8 into MySQL 5
- Yu Chengdong: Huawei is trying to cope with the US chip ban
- 天线的主要参数介绍
猜你喜欢

七个步骤,深入解读数据含义

深圳线下报名|StarRocks on AWS:如何对实时数仓进行极速统一分析

低码在数字化转型中扮演什么角色?

2023年网络安全预算规划的五个关键考虑因素

Leetcode--45. jumping game II (greed)

Personal production: ad library, component library, packaging library and 3D model, free of charge

Tcp/ip detailed diagram

The growth path of hardware engineers (0) -- Understanding components

Introduction to main parameters of antenna

连线:谁拥有未来的艺术?OpenAI允许Dall-E用户将作品商用化,目前而言
随机推荐
The diversion between video number and official account is convenient and considerable
solidity转账函数的实现(基于transfer)
Busybox latest version (busybox APK)
DC-DC开关电源
联发科已向美方提交申请,力求9月15日后仍可向华为供货
solidity的require报错
Outdoor activities in hot weather are exquisite! Here comes the safety guide for citizens' fitness in midsummer
Principle, classification and requirements of antenna
硬件工程师成长之路(0)----认识元件
The US ban on Huawei began to affect European chip manufacturers
Machine test questions for graduate students of Central South University
Go's walk library reports an error
Dormant for two years! Lingsheng technology's first aiot chip released: Samsung 11nm process, integrated NPU core!
天线的原理、分类及要求
Detailed explanation of network RJ45 interface
示波器参数详解
业绩增长强劲!闻泰科技上半年净利17亿元,同比暴增767.19%!
How to sharpen a knife simply by yourself
WordPress prompt error in establishing database connection
What role does low code play in the digital transformation?